On-site customized Gas Standard Preparation for single and multipoint calibration standards has never been easier than with the versatile GasMix? Gas Standard Preparation System.
The AlyTech GasMix? can prepare and automatically inject gases in virtually all kinds of analytical instruments. The principle behind GasMix? is based on mixing, diluting, and injecting two to twelve gas standards and may be used for a wide variety of applications like analyzer calibration and validation, to odorize natural gas, synthetic gas mixture generation, glove box atmosphere control,reactor gas feeding, sample conditioning prior to analysis.
● Ease of use
● Resourceful and transportable
● Reduced operator time
● Calibration curves for each gas used
●
Compliant with Internationalstandards method 6145time.
The GasMix? software is fully automated allowing for a single injection to run or a fully automated and pre- programmed sequence. The GasMix? Software can be remotely controlled; it may drive an external injection valve and/or the Start command of an analyzer. An added benefit of the Gasmix is drastically reduced operatorThe GasMix? software has a built in audit trail function that ensures a total traceability of the operations carried out.Regular automatic sequences may be saved and recalled at any time.
Environmental – Calibration of analyzers for ambient air surveillance in laboratories and on site
Petrochemical – Oil & Gas in laboratories
Food & Flavor - Quality control
Automotive Industry – Engine benchand emission control
Specialty gases
Electronics industry
? Number of channels: 2 to 6
Inlet gas pressure: 3 bar(g) to 8 bar(g) maximum
Outlet pressure: atmospheric up to 7 bar(g)
Connections: Swagelok 1/8“
Compatible gases: pure gases, concentrated mixtures, aggressive and reactive gases, with some exceptions (contact AlyTech or its representative for conffrmation)
Dilution factors up to 108
MFC Range: 2% to 100% of full scale.
Flow accuracy: ±1% of the required value over the entire scale.
Repeatability: less than 0.1% of the requested value
Software: Run on Windows 10 or higher. Ethernet port Communication
Operating temperature: 10 - 50°C
Alimentation: 90-260Vac, 1,5-4A, 50-60Hz
Dimension: L x l x h (cm): 45x22x38 (or 45x22x59 with options)
Approximate weight: 14 kg (depends on conffguration)